Chapter 8
Operating Instructions
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
8-9
Alarm
Silence
Function: Disables the audio alarm for two minutes. When Alarm Silence is
pressed before the end of a two-minute period, the two-minute timer is
reset. Alarms that cannot be silenced are listed in Table 9-1: “Alarm Alert
Messages” on page 9-5.
Indicator: Illuminated yellow when the audible alarm has been disabled; is
active and stays on for two minutes when the ALARM SILENCE button is
pressed. If ALARM SILENCE is active, and a new alarm condition occurs,
which involves exceeding an active alarm limit, the visual alarm functions
will be active. (Refer to “Alarm Silence” on page 9-2.) ALARM RESET
clears ALARM SILENCE. If a medium or high priority alarm exists after
ALARM RESET clears ALARM SILENCE, the audible alarm will begin.
Alarm
Reset
Function: Clears the visual indicator for auto reset alarms, certain active
alarms (see “Alarm Reset” on page 9-3), and reset of apnea ventilation
back to the active mode of ventilation (see “Apnea Ventilation” on page 8-
27). Alarm Reset also terminates ALARM SILENCE.
NOTE: The ventilator may automatically reset certain types of alarm
100% O
2
Function: Press once to deliver 100% O
2
to the patient for two minutes.
Subsequent button presses will reset the timer to two minutes.
NOTE: If the 100% O
alarm will be active for the two-minute 100%
Indicator: Illuminated green; is active only when the 100% O2 front panel
button has been pressed and the ventilator is delivering 100% O
2
to the
patient; remains on for the duration of 100% O
2
delivery (two minutes).
NOTE: If the operator sets the %O
indicator does not light. The 100% O
front panel key has been pressed..
Front Panel Keys (Continued)
Key Symbol
Definition
Description
Table 8-3: Front Panel Keys (Sheet 2 of 3)