Chapter 8
8-16
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Operating Instructions
Figure 8-14: Changing a Setting
When a value is entered, as shown in Figure 8-14, the ventilator checks to
assure that the value has been accepted by the operator, is within limits, and
will not cause other settings to be out of limits. If the new value causes these
limits to be exceeded, a diagnostic message will be displayed. (Refer to “Value
Entry Message” on page 12-4 for more details about the diagnostic messages).
Selecting Alarm Limits
Follow these steps to adjust currently active alarm limits:
1. Push ALARM SETTINGS. The ventilator displays the alarm limits for
the currently active ventilation mode (Figure 8-15).
2. Push the button for the alarm limit you want to adjust. The ventilator
displays the current value for that alarm limit in a display insert,
similar to Figure 8-14.
3. Press the bar or use the front panel knob to adjust the value.
4. Press ACCEPT to activate the new alarm limit or CANCEL to leave it
unchanged.
Current or new value
displayed digitally and
graphically
Press ACCEPT to
activate the new
setting
Press CANCEL to
leave the setting
unchanged
Push this bar to
decrease the value
Push this bar to
increase the value