Displayed data, Displayed data -111 – Philips V200 User Manual
Page 275

Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-111
Chapter 13
Displayed Data
While in Speaking Mode, the ventilator switches all volumetric readings from 
exhaled to delivered readings. This means that all data displayed based on 
exhaled volumes will now be based on delivered volumes. Some data is not 
available in Speaking Mode, including data on Patient Leak (circled below in 
Figure 9).
Figure 13-65: Active Mode: Displayed Data in the VCV Settings Screen
The following data in the Patient Data screen displays in delivered volumes 
(not exhaled volumes): 
•
Tidal Volume
•
Spontaneous Minute Volume
•
Total Minute Volume
The following data is not available in Speaking Mode. Dashed lines appear in 
the data box instead of a numeric value:
•
Patient Leak
•
F/Vt (Ratio of respiratory rate to tidal volume)
Patient Leak data 
is not available in 
Speaking Mode
