Rs-232 communications protocol – Philips V200 User Manual

Page 298

background image

Appendix A

Appendix A-12

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

RS-232 Communications Protocol

fields between the start and stop codes. The fourth field is the start code,
0x02. The last field in the string is the stop code, 0x03.

Volume Control Ventilation Settings Report

Description

Example

Resolution

Range

Units

Comments

Command Name

VCVS

N/A

N/A

N/A

Number of characters
between the start and
stop codes

209

N/A

N/A

N/A

3 character field

Number of fields
between the start and
stop codes

29

N/A

N/A

N/A

2 character field

Start Code

0x02

N/A

N/A

N/A

ASCII Start
Transmission
Character (STX)

Time of request

13:45

N/A

N/A

N/A

24 hour clock,
hh:mm

Date

FEB

◆23◆1997◆ N/A

N/A

N/A

12 character field,
MMM

◆DD◆YYYY◆

Current Ventilation
Type

VCV

◆◆◆

N/A

VCV

◆◆◆

PCV

◆◆◆

NPPV

◆◆

N/A

The ventilation
type currently
being used by the
ventilation.

VCV Mode Setting

A/C

◆◆◆

N/A

A/C

◆◆◆

SIMV

◆◆

CPAP

◆◆

N/A

VCV Waveform
Setting

RAMP

◆◆

N/A

RAMP

◆◆

SQUARE

N/A

VCV Patient Type

ADULT

N/A

ADULT

PED

◆◆◆

NEO

◆◆◆

VCV Respiratory Rate
Setting

12

◆◆◆◆

1

1 - 80

BPM

VCV Tidal Volume
Setting

500

◆◆◆

1

50 - 2500

mL

VCV Peak Flow
Setting

30

◆◆◆◆

1

3 - 140

LPM

VCV PEEP Setting

0

◆◆◆◆◆

1

0 - 35

cmH

2

O

(hPa)

VCV Pressure Support
Setting

0

◆◆◆◆◆

1

0 - 100

cmH

2

O

(hPa)

VCV I-Trigger Type

PRESS

N/A

PRESS

FLOW

◆◆

AUTO

◆◆

N/A

VCV Pressure I-
Trigger Setting

2.0

◆◆◆

0.1

0.1 - 20.0

cmH

2

O

(hPa)

Table A-2: Volume Control Ventilation Settings Report (Sheet 1 of 2)