Rs-232 communications protocol – Philips V200 User Manual
Page 298
Appendix A
Appendix A-12
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
RS-232 Communications Protocol
fields between the start and stop codes. The fourth field is the start code,
0x02. The last field in the string is the stop code, 0x03.
Volume Control Ventilation Settings Report
Description
Example
Resolution
Range
Units
Comments
Command Name
VCVS
N/A
N/A
N/A
Number of characters
between the start and
stop codes
209
N/A
N/A
N/A
3 character field
Number of fields
between the start and
stop codes
29
N/A
N/A
N/A
2 character field
Start Code
0x02
N/A
N/A
N/A
ASCII Start
Transmission
Character (STX)
Time of request
13:45
◆
N/A
N/A
N/A
24 hour clock,
hh:mm
◆
Date
FEB
◆23◆1997◆ N/A
N/A
N/A
12 character field,
MMM
◆DD◆YYYY◆
Current Ventilation
Type
VCV
◆◆◆
N/A
VCV
◆◆◆
PCV
◆◆◆
NPPV
◆◆
N/A
The ventilation
type currently
being used by the
ventilation.
VCV Mode Setting
A/C
◆◆◆
N/A
A/C
◆◆◆
SIMV
◆◆
CPAP
◆◆
N/A
VCV Waveform
Setting
RAMP
◆◆
N/A
RAMP
◆◆
SQUARE
N/A
VCV Patient Type
ADULT
◆
N/A
ADULT
◆
PED
◆◆◆
NEO
◆◆◆
VCV Respiratory Rate
Setting
12
◆◆◆◆
1
1 - 80
BPM
VCV Tidal Volume
Setting
500
◆◆◆
1
50 - 2500
mL
VCV Peak Flow
Setting
30
◆◆◆◆
1
3 - 140
LPM
VCV PEEP Setting
0
◆◆◆◆◆
1
0 - 35
cmH
2
O
(hPa)
VCV Pressure Support
Setting
0
◆◆◆◆◆
1
0 - 100
cmH
2
O
(hPa)
VCV I-Trigger Type
PRESS
◆
N/A
PRESS
◆
FLOW
◆◆
AUTO
◆◆
N/A
VCV Pressure I-
Trigger Setting
2.0
◆◆◆
0.1
0.1 - 20.0
cmH
2
O
(hPa)
Table A-2: Volume Control Ventilation Settings Report (Sheet 1 of 2)