Options and accessories – Philips V200 User Manual

Page 207

background image

Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-43

Chapter 13

Measured peak
inhalation pressure

0.0

0.1 or 1

0.0 – 99.9

100 – 130

cm
H

2

O

Outputs

when ventilation
mode is CPAP or
Spont and No
Apnea alarm is
present.

Measured mean
airway pressure

0.9

0.1 or 1

0.0 – 99.9

100 - 130

cm
H

2

O

Measured end
inhalation pressure

7.7

0.1 or 1

0.0 – 99.9

100 – 130

cm
H

2

O

Outputs

when

Ventilation type is
PCV or NPPV, or

Ventilation mode is
CPAP or Spont, or

VCV Insp. Hold
setting = 0.

Measured I:E ratio

0.2

0.1

0.1

9.9

10 - 99

N/A

Fractional
representation of
1:X.X with one (1)
decimal place,
e.g., 4:1 = 0.3

High Inhalation
Pressure Alarm
setting

20

1

10 – 105

cm
H

2

O

Outputs high
inhalation pressure
setting for active
ventilation type
(PCV or VCV) or
IPAP + 10 if NPPV
is active.

Low Inhalation
Pressure Alarm
setting

3

1

3 – 105

cm
H

2

O

Outputs Low
Inhalation
Pressure setting
for active
ventilation type.

Low PEEP or Low
EPAP Alarm setting

0

1

0

35

cm
H

2

O

Outputs Low PEEP
from active type
(VCV or PCV).
When active type
is NPPV uses Low
EPAP.

Ventilation Report

Description

Example

Resolution

Range

Units

Comments

Table 13-4: Ventilation Report (Sheet 4 of 10)