Rs-232 communications protocol – Philips V200 User Manual
Page 292
Appendix A
Appendix A-6
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
RS-232 Communications Protocol
NPPV High
Respiratory Rate
Alarm Limit Setting
150
◆◆◆
1
0 - 150
BPM
NPPV Low Minute
Volume Alarm Limit
Setting
1.00
◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0
L
NPPV Apnea Interval
Alarm Limit Setting
20
◆◆◆◆
1
10 - 60
sec
NPPV High Leak
Alarm Limit Setting
60
◆◆◆◆
1
10 - 60
sec
Measured Peak
Inspiratory Pressure
24.1
◆◆
0.1 for -20.0
to 99.9
1 for 100 to
130
-20.0 - 130 cmH
2
O
(hPa)
The PIP value from
the Patient Data
screen.
Measured Mean
Airway Pressure
5.6
◆◆◆
0.1 for -20.0
to 99.9
1 for 100 to
120
-20.0 -120
cmH
2
O
(hPa)
The MAP value
from the Patient
Data screen.
Measured End
Expiratory Pressure
2.0
◆◆◆
0.1
-20.0-99.9
cmH
2
O
(hPa)
The End Exp. value
from the Patient
Data screen.
Measured End
Inhalation Pressure
24.0
◆◆
0.1 for -20.0
to 99.9
1 for 100 to
130
-20.0 - 130 cmH
2
O
(hPa)
The Plateau value
from the Patient
Data screen.
Measured Tidal
Volume
468
◆◆◆
1
0 - 2500
mL
The Tidal Volume
value from the
Patient Data
screen.
Measured
Spontaneous Minute
Volume
0.00
◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0
L
The Spont VE value
from the Patient
Data screen.
Measured Minute
Volume
5.83
◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0
L
The Total VE value
from the Patient
Data screen.
Measured
Spontaneous Breath
Rate
0.0
◆◆◆
0.1 for 0.0
to 9.9
1 for 10 to
150
0.0 - 150
BPM
The Spont Rate
value from the
Patient Data
screen.
Measured Total Breath
Rate
12.0
◆◆
0.1 for 0.0
to 9.9
1 for 10 to
150
0 - 150
BPM
The Total Rate
value from the
Patient Data
screen.
Measured Rapid
Shallow Breathing
Index
5
◆◆◆◆◆
1
0 - 500
BPM/L
The F/Vt value from
the Patient Data
screen.
Ventilation Report (Continued)
Description
Example
Resolution
Range
Units
Comments
Table A-1: Ventilation Report (Sheet 5 of 10)