Rs-232 communications protocol – Philips V200 User Manual
Page 288
Appendix A
Appendix A-2
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
RS-232 Communications Protocol
Ventilation Report
Description
Example
Resolution
Range
Units
Comments
Command Name
VRPT
N/A
N/A
N/A
Number of characters
between the start and
stop codes
990
N/A
N/A
N/A
3 character field
Number of fields
between the start and
stop codes
134
N/A
N/A
N/A
3 character field
Start Code
0x02
N/A
N/A
N/A
ASCII Start
Transmission
Character (STX)
Time of request
13:45
◆
N/A
N/A
N/A
24 hour clock,
hh:mm
◆
Date
FEB
◆23◆1997◆ N/A
N/A
N/A
12 character field,
MMM
◆DD◆YYYY◆
Current Ventilation
Type
VCV
◆◆◆
N/A
VCV
◆◆◆
PCV
◆◆◆
NPPV
◆◆
N/A
The ventilation
type currently
being used by the
ventilator.
VCV Mode Setting
A/C
◆◆◆
N/A
A/C
◆◆◆
SIMV
◆◆
CPAP
◆◆
N/A
VCV Waveform Setting
RAMP
◆◆
N/A
RAMP
◆◆
SQUARE
N/A
VCV Patient Type
ADULT
◆
N/A
ADULT
◆
PED
◆◆◆
NEO
◆◆◆
VCV Respiratory Rate
Setting
12
◆◆◆◆
1
1 - 80
BPM
VCV Tidal Volume
Setting
500
◆◆◆
1
50 - 2500
mL
VCV Peak Flow Setting 30
◆◆◆◆
1
3 - 140
LPM
VCV PEEP Setting
0
◆◆◆◆◆
1
0 - 35
cmH
2
O
(hPa)
VCV Pressure Support
Setting
0
◆◆◆◆◆
1
0 - 100
cmH
2
O
(hPa)
VCV I-Trigger Type
PRESS
◆
N/A
PRESS
◆
FLOW
◆◆
AUTO
◆◆
N/A
VCV Pressure I-Trigger
Setting
2.0
◆◆◆
0.1
0.1 - 20.0
cmH
2
O
(hPa)
VCV Flow I-Trigger
Setting
3.0
◆◆◆
0.1
0.5 - 20.0
LPM
Table A-1: Ventilation Report (Sheet 1 of 10)