beautypg.com

Periodic maintenance, Periodic maintenance -7, Care and maintenance – Philips V200 User Manual

Page 129

background image

Chapter 10

Care and Maintenance

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

10-7

Periodic
Maintenance

This section includes detailed operator maintenance.

Schedule for Periodic Maintenance

Frequency

Component

Maintenance

During ventilator
setup

• Inspiratory bacteria filter

• Ventilator and patient circuit

components

• Inspiratory and expiratory filters

• Check filter for occlusions, cracks and

tears.

• Perform SST (Short Self-Test)

whenever circuit components are
changed

• Perform EST (Extended Self-Test)

between patient uses

• Ensure that the ventilator functions

normally with both filters in place

At least daily, and as
recommended by
filter manufacturers

• Inspiratory and expiratory filters

• Monitor performance of filters and

replace as needed. Review ventilator
patient graphics frequently for changes
in expiratory resistance which may
indicate degradation of expiratory filter.

• Follow filter manufacturer

recommendations regarding duration
of use, maintenance (for reusable
filters), removal and disposal. Note
that high humidity and aerosol
medications may reduce expiratory
filter life, increase expiratory
resistance, and/or cause filter damage.

CAUTION: Do not operate the ventilator without a properly functioning
expiratory filter and heater. Doing so may cause damage to delicate ventilator
components, such as the expiratory flow sensor, which may lead to inaccurate
spirometry or a Vent Inop condition.

WARNING: Vent Inop is a serious condition which is indicated by both visual and
audible alarms. If the ventilator is attached to a patient when Vent Inop occurs, the
patient must be supported with another means of life support ventilation.

At least daily

• Oxygen supply water trap and

filter

• Check and empty as required every

shift

At least every 250
hours

• Air Inlet & Fan Filters

• Inspect and clean. Some

environments cause a quicker
collection of lint and dust than others,
requiring maintenance more frequently
than every 250 hours.

Annually

• Annual preventive

maintenance kit (P/N
1034840). Kit contents are
subject to change.

• Install annual preventive maintenance

kit.

• Clean ventilator interior and exterior
• Complete performance verification

procedure

CAUTION: The annual preventive maintenance procedure is to be performed only
by a qualified service technician