Apnea ventilation, Technical specifications – Philips V200 User Manual
Page 147
Chapter 12
Technical Specifications
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
12-3
Non-Invasive Positive
Pressure Ventilation
Settings, Ranges and
Resolution
Resolution is one unit unless otherwise noted (cmH
2
O is considered
numerically equivalent to hPa).
Maximum delivered flow during pressure controlled breaths in NPPV is limited
to 100 LPM for the pediatric patient type.
Apnea Ventilation
Apnea ventilation is considered equivalent to Assist Control with Apnea Rate in
VCV and PCV. In NPPV, only mandatory breaths are delivered, triggered either
at the Apnea Rate or by the patient.
Rise Time
0.1 to 0.9 seconds
%O
2
21%
to
100%
Patient Type
Adult/Pediatric
Pressure Control Ventilation Settings, Ranges & Resolution (Continued)
Setting
Range
Table 12-4: Pressure Control Ventilation Settings, Ranges & Resolution (Sheet 2 of 2)
Non-Invasive Positive Pressure Ventilation Settings, Ranges & Resolution
Setting
Range
Respiratory Rate
1 to 80 Bpm
EPAP
2 to 25 cmH
2
O (hPa)
IPAP
2 to 35 cmH
2
O (hPa)
Inspiratory Time
0.1 to 9.9 seconds (Resolution is 0.1 second)
Rise Time
0.1 to 0.9 seconds
Inspiratory Trigger
(I-trigger)
Flow Sensitivity
0.5 to 20 LPM from base flow of 3 LPM above sensitivity (Resolution is
0.1 LPM)
Expiratory Trigger
(E-Cycle)
% Peak Flow
10 to 80% of inspiratory peak flow
%O
2
21%
to
100%
Apnea Rate
1-80 Bpm
Patient Type
Adult/Pediatric
Table 12-5: Non-Invasive Positive Pressure Ventilation Settings, Ranges and
Resolution