4-15
Battery Charging
When the ventilator is plugged into a viable AC supply, it will charge the
backup battery if the Mains Circuit Breaker is on and the machine is operating
as a ventilator, or in diagnostic mode or the Power ON/OFF switch is OFF
(
)
.
Charging time will depend on the amount of charge the batteries require. A
fully discharged backup battery will be fully recharged within 10 hours. If the
backup battery does not reach a full charge within 10 hours, contact
Respironics Customer Service. While the ventilator is charging, the front panel
indicator labeled “Charging” (text version of the front panel as shown in Figure
4-17) or the charging battery symbol
(symbol version of the front panel
as shown in Figure 4-18) will be on continuously. When the backup battery is
fully charged, the indicator will turn off.
NOTE:
NOTE:
Warranty
Respironics warrants the backup battery to be free from defects in material
and workmanship for a period of one year from the date of purchase, provided
that the unit is operated under conditions of normal use as described in this
operator’s manual.
At its discretion, Respironics will make replacements, repairs or issue credits
for equipment or parts that are found to be defective.