Options and accessories – Philips V200 User Manual
Page 212
Chapter 13
13-48
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
PCV Inhalation Time
Setting
0.10
0.01
0.10 – 9.90
sec
Outputs PCV
Inhalation Time
when PCV or VCV
is active, NPPV
Inhalation Time
when NPPV is
active
PCV or NPPV Apnea
Interval Setting
10
1
10 – 60
sec
Outputs PCV
Apnea Interval
when PCV or VCV
is active, NPPV
Apnea Interval
when NPPV is
active
PCV Apnea Inhalation
Pressure Setting or
NPPV Apnea IPAP
setting
100
1
5 – 100
cm
H
2
O
Outputs PCV
Pressure setting
when VCV or PCV
active, IPAP –
EPAP when NPPV
active
PCV Apnea
Respiratory Rate
Setting or NPPV
Apnea Respiratory
Rate Setting
12
0.1 or 1
1.0 - 9.9
10 – 80
BPM
Outputs PCV
Apnea Rate setting
when VCV or PCV
active, NPPV
Apnea rate when
NPPV active
PCV Apnea Inhalation
Time Setting or NPPV
Apnea Inhalation
Time Setting
2.00
0.01
0.10 – 9.90
sec
Outputs PCV
Inhalation Time
setting when VCV
or PCV active,
NPPV Inhalation
Time when NPPV
active
PCV Apnea Oxygen
Concentration Setting
or NPPV Apnea
Oxygen Concentration
100
1
21 – 100
%
PCV Apnea High
Inhalation Pressure
Setting or NPPV
Apnea High
Inhalation Pressure
Setting
20
1
10 – 105
cm
H
2
O
Outputs PCV High
Inhalation
Pressure setting
when VCV or PCV
active, IPAP + 10
when NPPV active
Alarm Silence Status
ON
N/A
ON
OFF
N/A
Apnea Alarm Status
NORMAL
N/A
NORMAL,
RESET
ALARM
N/A
Ventilation Report
Description
Example
Resolution
Range
Units
Comments
Table 13-4: Ventilation Report (Sheet 9 of 10)