Sterilization, Steam autoclaving, Chemical disinfecting – Philips V200 User Manual
Page 124: Sterilization -2, Care and maintenance
Chapter 10
10-2
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Care and Maintenance
Sterilization
Do not clean, disinfect, sterilize or reuse disposable products.
Steam Autoclaving
The O
2
sensor tee (P/N 1001736) and the O
2
sensor coupling
(P/N 1002505) may be steam autoclaved. The PVC O
2
sensor tee,
P/N 8-100498-00 (gray in color), cannot be autoclaved.
1. Ensure that the part has been disassembled, cleaned, and partially
reassembled, as applicable.
2. Separately wrap the part in muslin or equivalent paper wrapper.
3. Steam autoclave according to the autoclave manufacturers
instructions. In many institutions it may be routine to place a
biological indicator in the autoclave load as a subsequent test for
sterility.
4. Aseptically store parts until used.
Chemical Disinfecting
The O
2
sensor tee (P/N 1001736) and the O
2
sensor coupling
(P/N 1002505) may be chemically disinfected.The PVC O
2
sensor tee,
P/N 8-100498-00 (gray in color), cannot be chemically disinfected.
For chemically disinfected parts, reassemble after disinfecting.
WARNING:
Do not expose expiratory and inspiratory bacteria filters or reusable patient
tubing to ETO gas.
NOTE:
CAUTION:
CAUTION: