Percent leak, Percent leak -101, Options and accessories – Philips V200 User Manual

Page 265

background image

Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-101

Chapter 13

Percent Leak

When the neonatal software option is loaded on the ventilator, a % Leak
window appears on the Patient Data screen. The window displays the
estimated leak of the delivered volume as a percentage. The % Leak value is
updated at the beginning of each inhalation. The % Leak is available for all
patient types when the neonatal software option is loaded. While the estimate
for percent leak is not a precise measurement it can be used to trend relative
changes in leak delivered volume.

Figure 13-55: Patient Data Screen With % Leak

NOTE:

Patient data ranges and accuracies can be found in Chapter 12,
“Technical Specifications”
with the exception of the Exhaled Tidal
Volume. The accuracy for exhaled Tidal Volume is ±(4ml ±10%)

NOTE:

Neither Respiratory Mechanics nor Expiratory Hold are active when the
neonatal patient type is selected. Flow trigger is the only trigger type
available. All other features apply.

Percent Leak