Percent leak, Percent leak -101, Options and accessories – Philips V200 User Manual
Page 265
Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-101
Chapter 13
Percent Leak
When the neonatal software option is loaded on the ventilator, a % Leak
window appears on the Patient Data screen. The window displays the
estimated leak of the delivered volume as a percentage. The % Leak value is
updated at the beginning of each inhalation. The % Leak is available for all
patient types when the neonatal software option is loaded. While the estimate
for percent leak is not a precise measurement it can be used to trend relative
changes in leak delivered volume.
Figure 13-55: Patient Data Screen With % Leak
NOTE:
Patient data ranges and accuracies can be found in Chapter 12,
“Technical Specifications” with the exception of the Exhaled Tidal
Volume. The accuracy for exhaled Tidal Volume is ±(4ml ±10%)
NOTE:
Neither Respiratory Mechanics nor Expiratory Hold are active when the
neonatal patient type is selected. Flow trigger is the only trigger type
available. All other features apply.
Percent Leak