beautypg.com

Inspiratory area, Inspiratory area -24, Options and accessories – Philips V200 User Manual

Page 188

background image

Chapter 13

13-24

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

Options and Accessories

Figure 13-20: Loops Display with Overlay Feature Active

Inspiratory Area

The Loops screen includes a window for the inspiratory area (Figure 13-21F).
This window displays a numeric value that represents the ventilator imposed
work of breathing, and is only calculated and displayed for a frozen loop. If the
ventilator is supporting the entire breath, the Insp Area window displays
“0.000”. If the Insp Area window shows dashes (“– – – –”), the ventilator is
not calculating inspiratory area because the loops are not frozen.

Figure 13-21: Loops Screen with Inspiratory Area Window

SAVE button

OVERLAY icon
button

SINGLE VIEW
icon button

Dashes indicate
Inspiratory are
not calculated
(loop not frozen)