Rs-232 communications protocol – Philips V200 User Manual
Page 301
Appendix A
RS-232 Communications Protocol
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
Appendix A-15
PCV I-Trigger Type
PRESS
◆
N/A
PRESS
◆
FLOW
◆◆
AUTO
◆◆
N/A
PCV Pressure I-Trigger
Setting
2.0
◆◆◆
0.1
0.1 - 20.0
cmH
2
O
(hPa)
PCV Flow I-Trigger
Setting
3.0
◆◆◆
0.1
0.3 - 20.0
LPM
PCV E-Cycle Type
%FLOW
◆
N/A
%FLOW
◆
AUTO
◆◆
N/A
PCV %Flow Expiratory
Cycle Setting
25
◆◆◆◆
1
10 - 80
%
PCV Rise Time
0.1
◆◆◆
0.1
0.1 - 0.9
sec
PCV Oxygen
Concentration Setting
21
◆◆◆◆
1
21 - 100
%
PCV Apnea Rate
Setting
12
◆◆◆◆
1
1 - 80
BPM
PCV Apnea Interval
Alarm Limit Setting
20
◆◆◆◆
1
10 - 60
sec
PCV High Inspiratory
Pressure Alarm Limit
Setting
35
◆◆◆◆
1
10 - 105
cmH
2
O
(hPa)
PCV Low Inspiratory
Pressure Alarm Limit
Setting
3
◆◆◆◆◆
1
3 - 105
cmH
2
O
(hPa)
PCV Low PEEP Alarm
Limit Setting
0
◆◆◆◆◆
1
0 - 35
cmH
2
O
(hPa)
PCV Low Mandatory
Tidal Volume Alarm
Limit Setting
0
◆◆◆◆◆
1
0 - 2500
mL
PCV Low Spontaneous
Tidal Volume Alarm
Limit Setting
0
◆◆◆◆◆
1
0 - 2500
mL
PCV High Respiratory
Rate Alarm Limit
Setting
150
◆◆◆
1
0 - 150
BPM
PCV Low Minute
Volume Alarm Limit
Setting
1.00
◆◆
0.01 for
0.00 to
9.99
0.1 for
10.0 to
60.0
0.00 - 60.0
L
Pressure Control Ventilation Settings Report (Continued)
Description
Example
Resolution Range
Units
Comments
Table A-3: Pressure Control Ventilation Settings Report (Sheet 2 of 3)