Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-55

Chapter 13

5. At the end of the maneuver (whether successful or cancelled), the

Stop button changes to Continue. While the Continue button is
displayed, the graph is frozen and can be rescaled. Touch Continue or
exit the screen to unfreeze the display.

6. Touch Continue: the button changes to Start and you can repeat the

maneuver.

Figure 13-30: MIP/P0.1 Screen

WARNING:

Perform the MIP maneuver according to instructions. Because machine
assisted breath delivery, apnea detection, and detection of certain patient
alarms are temporarily disabled during a MIP maneuver, close clinical
supervision is recommended.

WARNING:

Because PEEP is temporarily set to 0 cmH

2

O

during a MIP maneuver, close

clinical supervision is recommended.

MIP
window

P0.1
window

Rescale
button

Start MIP
Press &
Hold
button

Scale
button

Start
P0.1
button