Extended self test (est), Extended self test (est) -12 – Philips V200 User Manual
Page 62
Chapter 5
5-12
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Setup
Press the OK button to clear the message from the screen. You may continue
to use the ventilator without the backup battery or power off the ventilator and
reconnect the backup battery.
To turn this feature off, enter Diagnostic Mode and deactivate the messaging
feature by pressing the BKUP BATTERY button.
1. Power off the ventilator.
2. Power on the ventilator while holding down the ALARM RESET and
100% O
2
keys simultaneously for approximately 5 seconds.
3. A message appears on the ventilator screen asking the user to ‘Verify
that the patient is disconnected prior to proceeding.’ Press OK to enter
Diagnostic Mode.
4. Once you’ve entered Diagnostic Mode, select USER CONFIG button,
which takes you to the User Config screen.
5. Press the BKUP BATTERY button. The background color should return
to blue signifying that this feature has been deactivated.
Extended Self Test
(EST)
We recommend that you run an Extended Self Test (EST) upon receipt of the
ventilator to ensure that there has been no shipping damage to the system. We
also recommend that you run EST between patients to verify the overall
functional integrity of the ventilator. Refer to chapter 11 for instructions on
running EST.
If EST was run successfully and all configuration information has been
entered, the ventilator is ready to be used. Follow the recommended
“Preoperational Procedure” on page 8-32 once a patient has been selected.
To exit the Diagnostics Mode you must turn the Power On/Off switch OFF (
)
and then turn it ON (
).
WARNING:
Always turn the ventilator power OFF before connecting additional
equipment.