beautypg.com

Testing, Battery specifications, Testing -11 battery specifications -11 – Philips V200 User Manual

Page 175: Options and accessories

background image

Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-11

Chapter 13

Testing

The external battery should be tested during the regular preventative
maintenance cycle of the ventilator or as required by your institutional
procedures. Perform the External Battery Verification Procedure included in
the External Battery Service Instructions (1013569).

Battery
Specifications

*Batteries are considered hazardous waste and must be disposed of according
to local regulations. It is unlawful to dispose of these batteries except through
a regionally approved recycling center.

For additional information, contact Respironics Customer Service at
1-800-345-6443.

CAUTION:

The External Battery is designed to recharge only as described here. Do
not use any other method to recharge the External Battery.

CAUTION:

If the ventilator is expected to be stored for more than 2 weeks, the
external battery ON/OFF switch should be set to the OFF (O) position to
avoid discharging the external battery. This switch must be placed in
the ON (I) position when the ventilator is removed from storage and
returned to use.

Part Number:

1059955

Model:

V200 External Battery

Type:

Lead Acid*

Output:

24 V, 33AH

Class:

1

Recharge Time:

~ 15 hours

Charge Hold:

~ 14 days at 20 C

Storage Conditions:

-20 C to 60 C at 10%-95% relative humidity