beautypg.com

Flow-trak, Option – Philips V200 User Manual

Page 247

background image

Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-83

Chapter 13

Flow-Trak

®

Option

Flow-Trak provides additional flow to Volume Control Ventilation (VCV) breaths.
It allows the patient to draw additional flow or volume as desired.

Ventilators with Flow-Trak installed are identified by the option label as seen in
Figure 13-39.

Figure 13-39: Flow-Trak option label

Once the Flow-Trak option has been loaded on the ventilator it is active. Follow
these steps to turn Flow-Trak Off and On:

1. While in VCV, press the Peak Flow settings button.

2. Press the Ftrak OFF button.

3. Press the Accept button.

Use these same steps pressing Ftrak On to turn Flow-Trak on.

Figure 13-40: Turning off the Flow-Trak Option

Flow-Trak will stay resident on the ventilator when shut down and restarted.
That is, if Flow-Trak is OFF when the ventilator is shut down then it will be OFF
the next time the ventilator is started.