Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-17

Chapter 13

5. When using oxygen from a wall supply, attach a wall oxygen hose (not

included) to the center inlet of the oxygen manifold assembly.

Replacement Parts:

The following replacement parts can be ordered from Respironics:

Check Valve Assembly (PN 1007190)

24-inch Medical Hose Assemblies - Oxygen (PN 1006655)

Bracket and Hardware Kit (PN 1062294)

Using the Manifold

To Use Wall Supply Oxygen

Connect the wall oxygen supply to the manifold. Close the O

2

cylinder valves.

Oxygen is delivered to the ventilator from all sources that are active. Therefore,
if the O

2

cylinder valves are open while oxygen is being delivered from the wall

supply, the cylinders can be depleted of oxygen. To avoid depleting the O

2

cylinders, shut off the cylinder valves when using the wall oxygen supply.

To Use O

2

Cylinders

Open the O

2

cylinder valves and disconnect the wall oxygen supply.

Transport Use

When preparing for intra-facility transport use, first open the O

2

cylinder valves

and then disconnect the wall oxygen supply. When transport is complete,
reconnect the wall oxygen supply and then close the O

2

cylinder valves.

WARNING:

If using wall oxygen supply, close the O

2

cylinder valves to avoid

depleting the cylinders.