Replacement parts, Using the manifold, To use wall supply oxygen – Philips V200 User Manual
Page 181: To use o2 cylinders, Transport use, Replacement parts: -17 using the manifold -17
Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-17
Chapter 13
5. When using oxygen from a wall supply, attach a wall oxygen hose (not
included) to the center inlet of the oxygen manifold assembly.
Replacement Parts:
The following replacement parts can be ordered from Respironics:
•
Check Valve Assembly (PN 1007190)
•
24-inch Medical Hose Assemblies - Oxygen (PN 1006655)
•
Bracket and Hardware Kit (PN 1062294)
Using the Manifold
To Use Wall Supply Oxygen
Connect the wall oxygen supply to the manifold. Close the O
2
cylinder valves.
Oxygen is delivered to the ventilator from all sources that are active. Therefore,
if the O
2
cylinder valves are open while oxygen is being delivered from the wall
supply, the cylinders can be depleted of oxygen. To avoid depleting the O
2
cylinders, shut off the cylinder valves when using the wall oxygen supply.
To Use O
2
Cylinders
Open the O
2
cylinder valves and disconnect the wall oxygen supply.
Transport Use
When preparing for intra-facility transport use, first open the O
2
cylinder valves
and then disconnect the wall oxygen supply. When transport is complete,
reconnect the wall oxygen supply and then close the O
2
cylinder valves.
WARNING:
If using wall oxygen supply, close the O
2
cylinder valves to avoid
depleting the cylinders.