Respironics V200 Ventilator Operator’s Manual

6-1

Chapter 6. Connecting Additional Equipment

Communication
Interface

The ventilator provides three communications interfaces: one serial RS-232
port, an analog output port, and a parallel port. (The parallel port is reserved
for use by the Communications Option.)

Figure 6-1: Ports and Outlets on the Rear Panel

WARNING:

Always turn the ventilator power OFF before connecting additional
equipment.

CAUTION:

To prevent the risk of excessive leakage due to external equipment being
connected to the ventilator via the communication ports, a means for
external separation of the conductive earth paths must be provided.

CAUTION:

All equipment used and connected to the ventilator communications
ports (analog, parallel, and serial) must comply with the medical
electrical equipment (IEC601-1) or other applicable standards.

Circuit Breaker
(cover not shown)

Humidifier AC
Outlet

AC Power Cord

Analog

Parallel

Remote Alarm

Serial