Technical specifications, D3.5 v, P = d 12 v – Philips V200 User Manual
Page 158: P = d 12 e, P = d 23 e

Chapter 12
12-14
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Technical Specifications
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ventilator is intended for use in the electromagnetic environment specified below. The user of 
the ventilator should assure that it is used in such an electromagnetic environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment - Guidance
Conducted 
RF
IEC 
61000-4-6
Portable and mobile RF communications 
equipment should be used no closer to any part of 
the ventilator, including cables, than the 
recommended separation distance calculated from 
the equation applicable to the frequency of the 
transmitter.Recommended separation distance 
Table 12-29: “EMC Declaration - Recommended 
Distances”.
3 Vrms150 
kHz to 80 
MHz outside 
ISM bands
a
3Vrms
Radiated 
RF
IEC 
61000-4-3
10 Vrms
150kHz to 
80 MHz in 
10 Vrms
10 V/m
80Mhz to 
2.5GHz
10 V/m
80Mhz to 800Mhz
800Mhz to 2.5GHz
where P is the maximum output power rating of the 
transmitter in watts (W) according to the 
transmitter manufacturer and d is the 
recommended separation distance in meters (m).
b
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey,
c
should be less than the compliance level in each 
frequency range.
d
Interference may occur in the vicinity of equipment 
marked with the 
following symbol: 
Table 12-28: EMC Declaration - Immunity
d
3.5
V
1
-------
P
=
d
12
V
2
------
P
=
d
12
E
1
------
P
=
d
23
E
1
------
P
=
