Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Technical Specifications
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ventilator is intended for use in the electromagnetic environment specified below. The user of
the ventilator should assure that it is used in such an electromagnetic environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment - Guidance
Conducted
RF
IEC
61000-4-6
Portable and mobile RF communications
equipment should be used no closer to any part of
the ventilator, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.Recommended separation distance
Table 12-29: “EMC Declaration - Recommended
Distances”.
3 Vrms150
kHz to 80
MHz outside
ISM bands
a
3Vrms
Radiated
RF
IEC
61000-4-3
10 Vrms
150kHz to
80 MHz in
10 Vrms
10 V/m
80Mhz to
2.5GHz
10 V/m
80Mhz to 800Mhz
800Mhz to 2.5GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
c
should be less than the compliance level in each
frequency range.
d
Interference may occur in the vicinity of equipment
marked with the
following symbol:
Table 12-28: EMC Declaration - Immunity
d
3.5
V
1
-------
P
=
d
12
V
2
------
P
=
d
12
E
1
------
P
=
d
23
E
1
------
P
=