Rs-232 communications protocol – Philips V200 User Manual

Page 290

background image

Appendix A

Appendix A-4

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

RS-232 Communications Protocol

PCV Inspiratory Time
Setting

1.00

◆◆

0.01

0.1 - 9.9

sec

PCV PEEP Setting

0

◆◆◆◆◆

1

0 - 35

cmH

2

O

(hPa)

PCV Pressure Support
Setting

0

◆◆◆◆◆

1

0 - 100

cmH

2

O

(hPa)

PCV I-Trigger Type

PRESS

N/A

PRESS

FLOW

◆◆

AUTO

◆◆

N/A

PCV Pressure I-Trigger
Setting

2.0

◆◆◆

0.1

0.1 - 20.0

cmH

2

O

(hPa)

PCV Flow I-Trigger
Setting

3.0

◆◆◆

0.1

0.3 - 20.0

LPM

PCV E-Cycle Type

%FLOW

N/A

%FLOW

AUTO

◆◆

N/A

PCV %Flow Expiratory
Cycle Setting

25

◆◆◆◆

1

10 - 80

%

PCV Rise Time

0.1

◆◆◆

0.1

0.1 - 0.9

sec

PCV Oxygen
Concentration Setting

21

◆◆◆◆

1

21 - 100

%

PCV Apnea Rate
Setting

12

◆◆◆◆

1

1 - 80

BPM

PCV Apnea Interval
Alarm Limit Setting

20

◆◆◆◆

1

10 - 60

sec

PCV High Inspiratory
Pressure Alarm Limit
Setting

35

◆◆◆◆

1

10 - 105

cmH

2

O

(hPa)

PCV Low Inspiratory
Pressure Alarm Limit
Setting

3

◆◆◆◆◆

1

3 - 105

cmH

2

O

(hPa)

PCV Low PEEP Alarm
Limit Setting

0

◆◆◆◆◆

1

0 - 35

cmH

2

O

(hPa)

PCV Low Mandatory
Tidal Volume Alarm
Limit Setting

0

◆◆◆◆◆

1

0 - 2500

mL

PCV Low Spontaneous
Tidal Volume Alarm
Limit Setting

0

◆◆◆◆◆

1

0 - 2500

mL

PCV High Respiratory
Rate Alarm Limit
Setting

150

◆◆◆

1

0 - 150

BPM

Ventilation Report (Continued)

Description

Example

Resolution

Range

Units

Comments

Table A-1: Ventilation Report (Sheet 3 of 10)