Chapter 10
Care and Maintenance
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
10-3
1. Separately immerse each part in the chemical disinfectant, i.e.
ammonia (15% solution), bleach (10% solution) or commercial,
hospital grade disinfectant, etc. Follow the disinfectant manufacturers
directions for the solution concentration, immersion times, and other
conditions for disinfecting. The use of a laminar air flow hood during
the chemical disinfecting process is recommended.
2. Thoroughly rinse and dry each part.
3. Aseptically reassemble (as required) and store the part until use.
Ventilator Exterior
Wipe the ventilator exterior clean with a damp cloth and mild detergent. Do not
use liquid or aerosol bactericide. Do not allow moisture to come in contact with
the touch panel screen or to collect between the keypad and front bezel
assembly.
Patient Circuit Support Arm
Wipe the patient circuit support arm clean with alcohol or bactericide.
O
2
Gas Supply Filter/Water Trap
Wipe the exterior of the O
2
gas supply filter/water trap with a mild solution of
soap and water. Rinse and dry parts. Do not steam-autoclave, chemically
disinfect, or expose to ETO gas. Do not allow liquid to migrate into the inlet
port.
CAUTION:
CAUTION:
CAUTION: