Chapter 12

Technical Specifications

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

12-9

Connectors

Alarm
Settings

High Inspiratory
Pressure

10 to 105 cmH

2

O (hPa)

Low Inspiratory
Pressure

3 to 105 cmH

2

O (hPa)

Low PEEP
Pressure

0 to 35 cmH

2

O (hPa)

High Respiratory
Rate

0 to 150 Bpm

Low Exhaled
Mandatory Tidal
Volume

0 mL to 2500 mL

Low Exhaled
Spontaneous Tidal
Volume

0 mL to 2500 mL

Low Exhaled
Minute Volume

0 to 60 Lpm

High Exhaled
Minute Volume

0 to 60 Lpm

High Leak (NPPV
mode only)

0 to 60 Lpm

Alarm
Status
Indicators

Normal Indicator

High Urgency Alarm (flashing red)

Medium/Low Urgency Alarm (flashing yellow/steady yellow)

Alarm Silence

Safety Valve Open

Ventilator Inoperative (steady red)

Screen Locked

Alarms (Continued)

Table 12-17: Alarms (Sheet 2 of 2)

Connectors

Inspiratory Limb Connector

Gas outlet port: 22 mm conical male

Expiratory Limb Connector

Gas return port (on expiratory filter):
Proprietary barb fitting

NOTE:

The gas return port on the ventilator is a
cylindrical port which requires mating to a
specified expiratory filter to seal the expiratory
limb.

Oxygen Inlet

DISS male, DISS female, NIST, Air Liquide, or SIS Fitting
(depending on country and configuration).

Table 12-18: Connectors