Page 196
Chapter 13
13-32
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
NOTE:
If "On/Off Preselect" is pressed a second time, the preselection is
switched off.
NOTE:
If there are already three devices preselected, one of the preselected
devices must be turned "off" before adding a new one.
7. Press "Store Preselect", then "Confirm".
8. Attach the adhesive label containing "Open Device" to the VueLink
module. The adhesive labels MUST be placed on the module in the
same order as they appear in the Preselection Task Window.
9. When the above steps have been completed, change the CMS to
Monitoring mode (see "Changing Modes").
For more information about the VueLink module, refer to the VueLink Module
Handbook.
Information Sent to the VueLink System
Once the CMS detects that the ventilator is connected to the module, the CMS
automatically displays ventilator parameters. The active type of ventilation
determines what data the CMS displays. Table Figure 13-24 defines CMS
displays and equivalent Respironics V200 Ventilator information.
Available
on
CareNet
CMS
message
V200
equivalent
Comments
Volume control ventilation (VCV) parameters
Mode and type of ventilation will always be displayed.
X
AWF
Flow Delivered
flow.
X
AWP
Pressure Delivered
pressure.
AWV
Volume
Delivered volume.
sPSV
set PSV
Set pressure support ventilation.
sPEEP
set PEEP
Set positive end expiratory pressure.
X
PIP
PIP
Measured peak inspiratory pressure.
MnAwP
MAP
Measured mean airway pressure.
Pplat
Pi End
Measured end inspiratory (plateau) pressure.
PEEP
Pe End
Measured positive end expiratory pressure.
SpMV
Spont VE
Measured spontaneous minute volume.
sTV
set TV
Set tidal volume.
MV
VE
Measured minute volume.
PtLeak
Pt Leak
Estimated patient leak.
X
TV
Vt
Measured tidal volume.
sFIO_2
set %O2
Set oxygen percentage (fractional inspired oxygen).