beautypg.com

Save and overlay features, Save and overlay features -23, Options and accessories – Philips V200 User Manual

Page 187

background image

Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-23

Chapter 13

Figure 13-19: Frozen Loops Display with Numeric Data Windows

Save and Overlay
Features

Follow these steps to save a frozen loop for later reference against future data
(for example, before and after bronchodilator therapy):

1. Press the FREEZE button to save the Loops screen.

2.

Use the arrow buttons or knob to select a breath.

3. Press the SAVE button.

4. Press CONTINUE to unfreeze the screen.

5. Press the OVERLAY button to superimpose a gray image of the saved loop

on the current display (the Overlay feature works on frozen or unfrozen
displays).

6. Press the SINGLE VIEW button to view current graphics without the saved

loops.

Figure 13-20 illustrates the Overlay feature

Cursor position
selects
waveform and
position (and
corresponding
points on loops)

Use arrow
buttons or knob
to set cursor
position.

Continue
button unfreeze
graphics