beautypg.com

Alarm status indicators, Operating instructions – Philips V200 User Manual

Page 83

background image

Chapter 8

Operating Instructions

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

8-5

Alarm Status Indicators

The alarm status indicators located at the top of the ventilator alert you to the
ventilator’s alarm conditions. (Refer to Chapter 9, “Alarms” for more detailed
descriptions.)

Figure 8-4: Alarm Status Indicators

Alarm Status Indicators

Alarm Indicator

Status

Description

Normal

Green

No active or auto reset alarm condition exists.

High

Flashing red

A high priority alarm condition exists.

Med/Low

Flashing
yellow

Continuous
yellow

A medium priority alarm condition exists.

A low priority alarm condition exists. In addition,
indicates auto reset conditions

Vent Inop

Red (active)

The ventilator is not capable of supporting
mechanical ventilation and requires service.
During Vent Inop, the ventilator opens the safety
valve to enable the patient to breathe room air
spontaneously. The ventilator also discontinues
detection of new alarm conditions during Vent
Inop.

WARNING: Vent Inop is a serious condition, which is

indicated by both visual and audible
alarms. If the ventilator is attached to a
patient when Vent Inop occurs, the
patient must be supported with another
means of life support ventilation.

Safety Valve

Red (active)

The safety valve is open and the ventilator is not in
operation. This is a high priority alarm condition.
When the Safety Valve indicator is illuminated, the
ventilator is not providing ventilatory support to
the patient. The safety valve opens to allow the
patient to breathe spontaneously through the
ventilator circuit. The patient must be capable of
creating a spontaneous breath in order to breathe
through the safety valve.

Table 8-1: Alarm Status Indicators

Normal

Alarm Med/Low

Alarm High

VENT
INOP

SAFETY
VALVE