beautypg.com

Rs-232 communications protocol – Philips V200 User Manual

Page 306

background image

Appendix A

Appendix A-20

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

RS-232 Communications Protocol

Dynamic
Resistance

5.43

◆◆

0.01 for 0.00 to
9.990.1 for 10.0 to
99.91 for 100 to 400

0.00 - 400

cmH2
0/L/
Sec

The Dynamic
Resistance display
from the
Mechanics Patient
Status Screen

Dynamic
Compliance

19.2

◆◆

0.01 for 0.00 to
9.990.1 for 10.0 to
99.91 for 100 to 350

0.00 - 350

mL/cm
H20

The Dynamic
Compliance display
from the
Mechanics Patient
Status Screen

Peak Lung Flow

35.1

◆◆

0.1 for -99.9 to 99.91
for -100 to -3001 for
100 to 300

-300 - 300

LPM

The Peak L-Flow
display from the
Mechanics Patient
Status Screen

Vital Capacity

468

◆◆

1

0 - 9999

mL

Vital Capacity
display from Vital
Capacity Screen

Vital Capacity Time
of Last Maneuver

11-14-01

◆◆13:11

1 minute

01-01-70
0:00 to

01-19-38

3:14

N/A

Time of Last
Maneuver display
from the Vital
Capacity
Screen.15
character field, 24
hour clock: MM-
DD-YY

uu

HH:MM

MIP

-54.1

0.1 for -99.9 to
99.91 for -100 and
100 to 200

-100 - 200

cm
H20

MIP display from
MIP/P0.1 display
Screen

MIP Time of Last
Maneuver

11-14-01

◆◆13:11

1 minute

01-01-7

0 0:00 to

01-19-38

3:14

N/A

MIP Time of Last
Maneuver display
from MIP/P0.1
Screen15
character field, 24
hour clock: MM-
DD-YY

uu

HH:MM

P0.1

-2.3

◆◆

0.1for -99.9 to
99.91 for -100 and
100 to 200

-100 - 200

cm
H20

P0.1 display from
MIP/P0.1 display
screen

P0.1 Time of Last
Maneuver

11-14-01

◆◆13:11

1 minute

01-01-70
0:00 to

01-1-/38

3:14

N/A

P0.1 Time of Last
Maneuver display
from MIP/P0.1
Screen.15
character field, 24
hour clock: MM-
DD-YYwwHH:MM

Static Resistance

5.43

◆◆

0,01 for 0.00 to
9.990.1 for 10.0 to
99.91 for 100 to
400

0.00 - 400

cmH2
0/L/
Sec

Static Resistance
display from the
Static C & R screen

Patient Data Report (Continued)

Description

Example

Resolution

Range

Units

Comments

Table A-5: Patient Data Report (Sheet 3 of 4)