Rs-232 communications protocol – Philips V200 User Manual

Page 291

background image

Appendix A

RS-232 Communications Protocol

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

Appendix A-5

PCV Low Minute
Volume Alarm Limit
Setting

1.00

◆◆

0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0

0.00 - 60.0

L

PCV High Minute
Volume Alarm Limit
Setting

60.0

◆◆

0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0

0.00 - 60.0

L

NPPV Mode Setting

SPONT/T

N/A

SPONT/T

SPONT

◆◆

NPPV Patient Type

ADULT

N/A

ADULT

PED

◆◆◆

NEO

◆◆◆

NPPV Respiratory
Rate Setting

12

◆◆◆◆

1

1 - 80

BPM

NPPV EPAP Setting

5

◆◆◆◆◆

1

2 - 25

cmH

2

O

(hPa)

NPPV IPAP Setting

5

◆◆◆◆◆

1

2 - 35

cmH

2

O

(hPa)

NPPV Inspiratory Time
Setting

1.0

◆◆◆

0.1

0.1 - 9.9

sec

NPPV Rise Time

0.1

◆◆◆

0.1

0.1 - 0.9

sec

NPPV I-Trigger Type

AUTO

N/A

FLOW

◆◆

AUTO

◆◆

N/A

NPPV Flow I-Trigger
Setting

3.0

◆◆◆

0.1

0.5 - 20.0

LPM

NPPV E-Cycle Type

AUTO

◆◆

N/A

FLOW

◆◆

AUTO

◆◆

N/A

NPPV %Flow
Expiratory Cycle
Setting

25

◆◆◆◆

1

10 - 80

%

NPPV Oxygen
Concentration Setting

21

◆◆◆◆

1

21 - 100

%

NPPV Apnea Rate
Setting

12

◆◆◆◆

1

1 - 80

BPM

NPPV Low Inspiratory
Pressure Alarm Limit
Setting

3

◆◆◆◆◆

1

3 - 105

cmH

2

O

(hPa)

NPPV Low EPAP
Alarm Limit Setting

0

◆◆◆◆◆

1

0 - 25

cmH

2

O

(hPa)

NPPV Low Tidal
Volume Alarm Limit
Setting

0

◆◆◆◆◆

1

0 - 2500

mL

Ventilation Report (Continued)

Description

Example

Resolution

Range

Units

Comments

Table A-1: Ventilation Report (Sheet 4 of 10)