Rs-232 communications protocol – Philips V200 User Manual
Page 291
Appendix A
RS-232 Communications Protocol
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
Appendix A-5
PCV Low Minute
Volume Alarm Limit
Setting
1.00
◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0
L
PCV High Minute
Volume Alarm Limit
Setting
60.0
◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0
L
NPPV Mode Setting
SPONT/T
N/A
SPONT/T
SPONT
◆◆
NPPV Patient Type
ADULT
◆
N/A
ADULT
◆
PED
◆◆◆
NEO
◆◆◆
NPPV Respiratory
Rate Setting
12
◆◆◆◆
1
1 - 80
BPM
NPPV EPAP Setting
5
◆◆◆◆◆
1
2 - 25
cmH
2
O
(hPa)
NPPV IPAP Setting
5
◆◆◆◆◆
1
2 - 35
cmH
2
O
(hPa)
NPPV Inspiratory Time
Setting
1.0
◆◆◆
0.1
0.1 - 9.9
sec
NPPV Rise Time
0.1
◆◆◆
0.1
0.1 - 0.9
sec
NPPV I-Trigger Type
AUTO
◆
N/A
FLOW
◆◆
AUTO
◆◆
N/A
NPPV Flow I-Trigger
Setting
3.0
◆◆◆
0.1
0.5 - 20.0
LPM
NPPV E-Cycle Type
AUTO
◆◆
N/A
FLOW
◆◆
AUTO
◆◆
N/A
NPPV %Flow
Expiratory Cycle
Setting
25
◆◆◆◆
1
10 - 80
%
NPPV Oxygen
Concentration Setting
21
◆◆◆◆
1
21 - 100
%
NPPV Apnea Rate
Setting
12
◆◆◆◆
1
1 - 80
BPM
NPPV Low Inspiratory
Pressure Alarm Limit
Setting
3
◆◆◆◆◆
1
3 - 105
cmH
2
O
(hPa)
NPPV Low EPAP
Alarm Limit Setting
0
◆◆◆◆◆
1
0 - 25
cmH
2
O
(hPa)
NPPV Low Tidal
Volume Alarm Limit
Setting
0
◆◆◆◆◆
1
0 - 2500
mL
Ventilation Report (Continued)
Description
Example
Resolution
Range
Units
Comments
Table A-1: Ventilation Report (Sheet 4 of 10)