Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-57

Chapter 13

3. At the end of a successful maneuver, a pop-up screen displays the

result of the maneuver and asks you to accept or reject the results by
touching the Accept or Reject button.

4. At the end of the maneuver (whether successful or cancelled), the

Stop button changes to Continue. While the Continue button is
displayed, the graph is frozen and can be rescaled. Touch Continue or
exit the screen to unfreeze the display.

5. Touch Continue: the button changes to Start and you can repeat the

maneuver.

6. Return the active ventilation mode from VCV to the previous mode if

necessary.

Figure 13-31: Static C and R Screen

Information type

Description

Pressure--Volume or
Pressure--Time button

Pressure--Volume: Displays a real-time pressure-volume loop.
Pressure--Time: Displays a real-time pressure-time waveform.

Start button

Starts the maneuver. Changes to Stop when the maneuver is started.
Changes to Continue when the maneuver is complete or when the Stop
button is touched. Changes to Start when Continue is touched.

Static C and Static R
windows

Displays the most recently accepted maneuver results. The date and
time of the maneuver appear below each window.

Table 13-7: Summary of Static C and R Screen Information

Time
Stamp

Rescale
button

Start
button

Static C and R Windows

Pressure-Volume and Pressure-Time buttons

P Plat
window

Scale