beautypg.com

Installation, Power consumption sequence, External battery/ backup battery operation – Philips V200 User Manual

Page 172

background image

Chapter 13

13-8

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

Options and Accessories

Installation

Contact Respironics Customer Service at 1-800-345-6443 to arrange
installation or refer to the installation section of the Ventilator External Battery
Service Instructions (1013569).

Power Consumption
Sequence

When AC power becomes unavailable, the ventilator consumes power from the
battery systems as follows:

With the external battery installed and turned ON and backup battery
installed:
External battery power is consumed first until depleted.
Backup battery power is then used until depleted. An audible alarm
sounds when the backup battery is in use. The ventilator will not
operate when both batteries are depleted.

With backup battery installed only: Backup battery power is used until
depleted. An audible alarm sounds when the backup battery is in use.
The ventilator will not operate when the battery is depleted.

AC power can be restored to the ventilator at any time during battery use.
When AC power is restored the ventilator consumes AC power only, and both
batteries begin recharging. (See “Battery Charging” on page 13-10).

External Battery/
Backup Battery
Operation

The external battery must be turned ON in order for the external battery to be
used. Make sure the ON/OFF switch located on the external battery (Figure 13-
6) is in the ON (I)
position. When AC power becomes unavailable, the
ventilator automatically switches to external battery operation without
interruption in ventilation. When the external battery is depleted, the ventilator
automatically switches to backup battery power.

WARNING:

Do not lift the external battery. If the battery needs to be removed from the
cart, contact Respironics Customer Service at 1-800-345-6443.