Technical Specifications

Value Entry Message

If you enter a value that is outside the operational limits of the ventilator, you
will be shown a message window on the screen. Once you acknowledge the
message, the ventilator returns to the value modification display without
changing the value.

Assist/Control Ventilation

Value Entry Reasonability Checks

Item

Limit Exceeded

Message

Apnea Rate

I:E Ratio>3:1

I:E Ratio must be less than or equal to 3:1.
Check V

t

, Peak Flow, or Insp. Hold

Peak Flow

I:E Ratio>3:1

I:E Ratio must be less than or equal to 3:1.
Check V

t

, Apnea Rate, or Insp. Hold.

Tidal Volume

I:E Ratio>3:1

I:E Ratio must be less than or equal to 3:1.
Check Flow, Apnea Rate, or Insp. Hold.

Insp. Hold

I:E Ratio>3:1

I:E Ratio must be less than or equal to 3:1.
Check V

t

, Apnea Rate or Peak Flow.

Waveform

I:E Ratio>3:1

I:E Ratio must be less than or equal to 3:1.
Check V

t

, Peak Flow or Apnea Rate.

I-Time

I-Time>9 sec.

I-Time must be less than or equal to 9 sec.
Check V

t

, Peak Flow, Insp. Hold or Waveform.

Table 12-6: Assist/Control Ventilation Value Entry Reasonability Checks

Pressure Control Ventilation

Value Entry Reasonability Checks

Item

Limit Exceeded

Message

Changed

Apnea Rate

I:E Ratio>4:1

I:E Ratio must be less than
or equal to 4:1.Check I-
Time.

Readjust the Apnea Rate or
adjust I-Time or Rate.

I-Time

I:E Ratio>4:1

I:E Ratio must be less than
or equal to 4:1.Check
Apnea Rate.

Readjust the I-Time or
adjust Apnea Rate or Rate.

I-Time

I-Time

I-Time must greater than or
equal to Rise Time.

Increase the Rise Time or
increase the I-Time.

Rise Time

Rise Time>I Time

Rise Time must be less
than I-Time.

Decrease the Rise Time or
increase the I-Time.

Table 12-7: Pressure Control Ventilation Value Entry Reasonability Checks