Options and accessories – Philips V200 User Manual

Page 223

background image

Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-59

Chapter 13

Maneuver

Error Messages

Temporarily Disabled Alarms

VC

Measurement out of range
Measured volume exceeds 9999 mL.

Vital Capacity unavailable due to alarm
condition
Alarm condition exists when maneuver
requested.

Maneuver cancelled
Operator presses Stop button.

Maneuver timed out
No inspiration within 20 seconds of starting
maneuver.

Maneuver cancelled by alarm condition
Alarm condition occurs after maneuver
begins.

•Low Exhaled Minute Volume

•Low Exhaled Mandatory Tidal

Volume

•High Exhaled Minute Volume

•Low Exhaled Spontaneous Tidal

Volume

•Low PEEP

MIP

Measurement out of range
Measured pressures exceed –100 to +200
cmH

2

O.

MIP unavailable due to alarm condition
Alarm condition exists when maneuver
requested.

Maneuver timed out
MIP button held down for 30 seconds since
starting maneuver.

All patient alarms and the Low O2
alarm are disabled during the MIP
maneuver.

Table 13-8: Respiratory Mechanics Error Messages and Alarms