Chapter 8
Operating Instructions
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
8-33
7. Push USER CONFIG to check time, time format, date and altitude and
set compliance compensation as required (Figure 8-37).
8. Press START SST to begin Short Self Test, and follow the screen
prompts. Use a cap to plug the patient port of the patient circuit wye
at the prompt.
9. During SST, as each test is performed, the ventilator displays test
results (Figure 8-37). Do not proceed until the ventilator completes
SST without failures.
Figure 8-37: SST Results
NOTE:
Any failures
detected during
SST are listed
here.
Push USER
CONFIG to set
time, date,
altitude,
compliance
compensation
enable, and time
format.
WARNING:
You will be warned if the compliance is 9.0 ml/cmH
Patients should not be put on a patient circuit that does not meet this
requirement.