Patient data screen, Patient data screen -5, Technical specifications – Philips V200 User Manual
Page 149
Chapter 12
Technical Specifications
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
12-5
Patient Data Screen
Monitored patient data is displayed when Patient Data is pressed. Patient data
ranges, resolution, units, and accuracy specifications are provided in the table
below. Minute volume data is based on an eight breath average.
Non-Invasive Ventilation
Value Entry Reasonability Checks
Item
Limit Exceeded
Message
Changed
Apnea Rate
I:E Ratio>4:1
I:E Ratio must be less than
or equal to 4:1. Check I-
Time.
Readjust the Apnea Rate or
adjust I-Time or Rate.
I-Time
I:E Ratio>4:1
I:E Ratio must be less than
or equal to 4:1. Check
Apnea Rate.
Readjust the I-Time or
adjust Apnea Rate or Rate.
IPAP
IPAP IPAP must be greater than Readjust IPAP higher or EPAP EPAP>IPAP EPAP must be less than or Readjust EPAP lower or Rise Time Rise Time>I Time Rise Time must be less Decrease the Rise Time or Table 12-8: Non-Invasive Ventilation Value Entry Reasonability Checks General Value Entry Reasonability Checks Item Limit Exceeded Message Changed Apnea Apnea Rate greater than or equal to the Increase the Apnea Rate or Table 12-9: General Value Entry Reasonability Checks Patient Data Range, Resolution, Units & Accuracy Patient Data Display Range Units Resolution Accuracy Exhaled Minute 0.00 to 99.9 L 0.01 for 0.00-9.99; ±10% Exhaled Tidal Volume 0 to 9999 mL 1 ±10% Spontaneous Minute 0.00 to 99.9 L 0.01 for 0.00-9.99; ±10% Rapid Shallow Breathing 0 to 500 Bpm/L 1 ±10% Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 1 of 2)
or equal to EPAP.
adjust EPAP lower.
equal to IPAP.
adjust IPAP higher.
than I-Time.
increase the I-Time.
Set Rate.
decrease the Rate.
Volume
0.1 for 10.0-99.9
Volume
0.1 for 10.0-99.9
Index