Patient Data Screen

Monitored patient data is displayed when Patient Data is pressed. Patient data
ranges, resolution, units, and accuracy specifications are provided in the table
below. Minute volume data is based on an eight breath average.

Non-Invasive Ventilation

Value Entry Reasonability Checks

Item

Limit Exceeded

Message

Changed

Apnea Rate

I:E Ratio>4:1

I:E Ratio must be less than
or equal to 4:1. Check I-
Time.

Readjust the Apnea Rate or
adjust I-Time or Rate.

I-Time

I:E Ratio>4:1

I:E Ratio must be less than
or equal to 4:1. Check
Apnea Rate.

Readjust the I-Time or
adjust Apnea Rate or Rate.

IPAP

IPAP

IPAP must be greater than
or equal to EPAP.

Readjust IPAP higher or
adjust EPAP lower.

EPAP

EPAP>IPAP

EPAP must be less than or
equal to IPAP.

Readjust EPAP lower or
adjust IPAP higher.

Rise Time

Rise Time>I Time

Rise Time must be less
than I-Time.

Decrease the Rise Time or
increase the I-Time.

Table 12-8: Non-Invasive Ventilation Value Entry Reasonability Checks

General Value Entry Reasonability Checks

Item

Limit Exceeded

Message

Changed

Apnea

Apnea Rate

greater than or equal to the
Set Rate.

Increase the Apnea Rate or
decrease the Rate.

Table 12-9: General Value Entry Reasonability Checks

Patient Data Range, Resolution, Units & Accuracy

Patient Data

Display Range

Units

Resolution

Accuracy

Exhaled Minute
Volume

0.00 to 99.9

L

0.01 for 0.00-9.99;
0.1 for 10.0-99.9

±10%

Exhaled Tidal Volume

0 to 9999

mL

1

±10%

Spontaneous Minute
Volume

0.00 to 99.9

L

0.01 for 0.00-9.99;
0.1 for 10.0-99.9

±10%

Rapid Shallow Breathing
Index

0 to 500

Bpm/L

1

±10%

Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 1 of 2)