Chapter 13

13-100

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

Options and Accessories

Initial Neonatal Settings

When the ventilator is first powered on and the neonatal patient type is
selected, the initial startup settings found in Table 13-21: are applied. Each
subsequent startup will display the settings from the last use of the ventilator
of that patient type with the exception of O

2

. The last O

2

setting used on the

ventilator will remain regardless of patient type.

Initial Patient Settings, Ranges

Ventilation Settings

Value

Ranges

Ventilation Mode

PCV - SIMV

AC, SIMV, CPAP

Rate

20 BPM

1 - 150 BPM

Pressure

10 cmH

2

O

5 - 100 cmH

2

O

I-Time

0.3 Sec

0.1 - 2.0 Sec

PEEP

3 cmH

2

O

0 - 35 cmH

2

O

PSV

0 cmH

2

O

0 - 100 cmH

2

O

I-Trigger

2.0 L/Min

0.3 - 10.0 L/Min

E-Cycle

25%

10 - 80%

Rise-TIme

0.2 Sec

0.1 - 0.5 Sec

O

2

%

same as last O

2

setting

21 - 100%

Apnea Rate

20 BPM

1 - 80 BPM

Alarm Settings

Value

Ranges

High Press

30 cmH2O

10 - 105 cmH2O

Low Insp Press

5 cmH2O

3 - 105 cmH2O

Low Peep

0 cmH2O

0 - 35 cmH2O

Low Vt Mand

10 mL

0 - 300 mL

Low Vt Spont

3 mL

0 - 300 mL

High Rate

150 BPM

0 - 150 BPM

High VE

1.2 L

0.00 - 5.0 L

Low VE

0.3 L

0.00 - 5.0 L

Apnea

20 Sec

10 - 60 Sec

I-Time Too Long

sounds after 1.5 Sec of spont
breath

Table 13-21: Initial Neonatal Patient Settings