Operating Instructions

Patient Data Screen

You can view the patient data screen (Figure 8-33) by pressing the PATIENT
DATA button in the top bar. The format of this screen is the same in VCV, PCV,
and NPPV. Table 8-7: “Patient Data Definitions: Range, Units & Resolution”
summarizes the definitions, display ranges, and resolution for patient data
parameters.

Figure 8-33: Patient Data Screen

Patient Data Definitions: Range, Units & Resolution

Parameter

Description

Display Range

Resolution

PIP

Peak Inspiratory Pressure
The maximum airway pressure
during inspiration.

-20.0 to 130
cmH

2

O (hPa)

0.1 for -20.0 to
99.9
1 for 100 to 130

MAP

Mean Airway Pressure
The average of airway pressure
during one full breath cycle.

-20.0 to 120
cmH

2

O (hPa)

0.1 for -20.0 to
99.9
1 for 100 to 130

Pe End

Pressure at End Expiration
The pressure measured at the
end of expiration.

-20.0 to 99.9
cmH

2

O (hPa)

0.1

Pi End

Pressure at End Inspiration
The pressure measured at the
end of inspiration (or at the end
of Insp Hold if it is >0).

-20.0 to 130
cmH

2

O (hPa)

0.1 for -20.0 to
99.9
1 for 100 to 130

Tidal Vol.

Exhaled Tidal Volume
The volume exhaled at each
breath. Compliance
compensated (if enabled).

0 to 9999 ml
(BTPS)

1

Table 8-7: Patient Data Definitions: Range, Units & Resolution (Sheet 1 of 2)