Options and accessories – Philips V200 User Manual
Page 217
Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-53
Chapter 13
3. At the end of a successful maneuver, a pop-up screen displays the
result of the maneuver and asks you to accept or reject the results by
touching the Accept or Reject button.
4. At the end of the maneuver (whether successful or cancelled), the
Stop button changes to Continue. While the Continue button is
displayed, the graph is frozen and can be rescaled. Touch Continue or
exit the screen to unfreeze the display.
5. Touch Continue: the button changes to Start and you can repeat the
maneuver.
Figure 13-29: Vital Capacity Screen
WARNING:
Perform the Vital Capacity maneuver according to instructions. Because
machine-assisted breath delivery, apnea detection, and detection of certain
patient alarms are temporarily disabled during a Vital Capacity maneuver,
close clinical supervision is recommended.
WARNING:
Because PEEP is temporarily set to 0 cmH
2
O
during a Vital Capacity
maneuver, close clinical supervision is recommended.
Flow-Volume and Volume-Time buttons
Time
Stamp
Rescale
button
Vital
Capacity
window
Start
button
Start
button
Vital
Capacity
window
Flow-Volume and Volume-Time buttons
Scale
button