Options and accessories – Philips V200 User Manual

Page 217

background image

Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-53

Chapter 13

3. At the end of a successful maneuver, a pop-up screen displays the

result of the maneuver and asks you to accept or reject the results by
touching the Accept or Reject button.

4. At the end of the maneuver (whether successful or cancelled), the

Stop button changes to Continue. While the Continue button is
displayed, the graph is frozen and can be rescaled. Touch Continue or
exit the screen to unfreeze the display.

5. Touch Continue: the button changes to Start and you can repeat the

maneuver.

Figure 13-29: Vital Capacity Screen

WARNING:

Perform the Vital Capacity maneuver according to instructions. Because
machine-assisted breath delivery, apnea detection, and detection of certain
patient alarms are temporarily disabled during a Vital Capacity maneuver,
close clinical supervision is recommended.

WARNING:

Because PEEP is temporarily set to 0 cmH

2

O

during a Vital Capacity

maneuver, close clinical supervision is recommended.

Flow-Volume and Volume-Time buttons

Time
Stamp

Rescale
button

Vital
Capacity
window

Start
button

Start
button

Vital
Capacity
window

Flow-Volume and Volume-Time buttons

Scale
button