Chapter 4

Getting Started

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

4-21

Oxygen Source Connection Instructions

Refer to Figure 4-21.

1. Locate the O

2

filter with water trap (1) located in the lower left corner

of the rear panel.

2. Ensure that the O

2

filter and water trap (1) are properly attached to

the ventilator. Inspect the bowl (1) for cracks or potential leaks.
Ensure all connections are tight.

3. Inspect the O

2

hose (2), hose connector (3) and the hose connector

mate (4).

4. Check the oxygen gas supply. Clean, adjust pressure, and drain

condensate from any water traps, filters, or regulators in the O

2

supply

lines.

5. Connect O

2

gas connector (3) to the ventilator O

2

filter with DISS hose

connector (4).

6. If gas source is not already turned on, turn it on and verify that no O

2

gas is leaking.