beautypg.com

Removing and replacing o2 input filter element, Care and maintenance – Philips V200 User Manual

Page 131

background image

Chapter 10

Care and Maintenance

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

10-9

Removing and Replacing O

2

Input Filter Element

Figure 10-5: Removing and Replacing O

2

Input Filter Element

Refer to Figure 10-5.

1. Disconnect O

2

Hose (1).

2. Unscrew filter housing (3).

3. Unscrew filter element retaining screw (4).

4. Remove and replace filter element (5).

5. Reconnect filter element retaining screw (4).

6. Reinstall filter housing (3).

7. Reconnect O

2

hose (1).

8. Check system for leaks prior to patient use.

1

5

2

3

4