Removing and replacing o2 input filter element, Care and maintenance – Philips V200 User Manual
Page 131
Chapter 10
Care and Maintenance
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
10-9
Removing and Replacing O
2
Input Filter Element
Figure 10-5: Removing and Replacing O
2
Input Filter Element
1. Disconnect O
2
Hose (1).
2. Unscrew filter housing (3).
3. Unscrew filter element retaining screw (4).
4. Remove and replace filter element (5).
5. Reconnect filter element retaining screw (4).
6. Reinstall filter housing (3).
7. Reconnect O
2
hose (1).
8. Check system for leaks prior to patient use.
1
5
2
3