User configuration screen, Date and time, Setting date and time format – Philips V200 User Manual
Page 57: User configuration screen -7
Chapter 5
Setup
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
5-7
User Configuration
Screen
The buttons across the top and bottom bars are used to select the Diagnostic
Mode functions. To set up the machine for the first time for operation as a
ventilator, USER CONFIG should be selected. When this button is pressed, the
following screen appears.
Figure 5-6: User Configuration Screen
The User Configuration screen allows the operator to:
•
Set the date and time
•
Activate or deactivate the automatic patient circuit compliance
compensation feature
•
Set the proper altitude
•
Set 24 Hrs/AM PM time display
•
Activate or deactivate backup battery check at startup
Date and Time
The real-time clock and calendar will last for approximately 2.5 years. When
the ventilator is received, the clock’s time will have to be changed to that of
the existing time zone. The date should be checked and changed if necessary.
Setting Date and Time Format
To set time (Figure 5-8) or date (Figure 5-7):
1. Press the desired date or time button (month, day, year, hour, minute,
second).
2. When the window insert appears, press the INCREASE or DECREASE
button, or rotate the control knob to change the value of the selected
parameter.