Filters, Measuring and display devices, Ac power and battery indicators – Philips V200 User Manual
Page 154: Technical specifications
Chapter 12
12-10
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Technical Specifications
Filters
Measuring and
Display Devices
AC Power and Battery
Indicators
Filters
Inspiratory
See filter instruction sheets for complete specification.
Bacterial Filter Efficiency: 99.999+%
Viral Filtration Efficiency: 99.99+%
Expiratory
See filter instruction sheets for complete specification.
Typical efficiency: 99.97% for nominal particle size of 0.3 micro meter (micron)
at 100L/min flow.
Table 12-19: Filters
Measuring and Display Devices
Pressure Measurements
Type
Solid State
Sensing Position
Inhalation output port
Exhalation input port
Measurements
Patient airway pressure
Volume Measurements
Type
Hot Film Anemometer flow integrated
Sensing Position
O
2
, air and exhalation
Measurements
O
2
flow, air flow and exhalation flow
Oxygen Measurements
Type
Zirconia solid electrolyte
Sensing Position
Internal enclosure
Measurements
Oxygen concentration (range 0 to 95%)
Type
Galvanic cell
Sensing Position
Inspiratory limb of VBS
Measurements
Delivered O
2
% (range 0 to 110%)
Settings, Alarms and
Data Display
Type
TFT liquid crystal touch screen
Table 12-20: Measuring and Display Devices
AC Power & Battery Indicators
Loss of Power (audible only)
Backup Battery Low
External Battery In Use
Backup Battery In Use
Backup Battery Charging
Mains