Technical Specifications

Filters

Measuring and
Display Devices

AC Power and Battery
Indicators

Filters

Inspiratory

See filter instruction sheets for complete specification.

Bacterial Filter Efficiency: 99.999+%
Viral Filtration Efficiency: 99.99+%

Expiratory

See filter instruction sheets for complete specification.

Typical efficiency: 99.97% for nominal particle size of 0.3 micro meter (micron)
at 100L/min flow.

Table 12-19: Filters

Measuring and Display Devices

Pressure Measurements

Type

Solid State

Sensing Position

Inhalation output port
Exhalation input port

Measurements

Patient airway pressure

Volume Measurements

Type

Hot Film Anemometer flow integrated

Sensing Position

O

2

, air and exhalation

Measurements

O

2

flow, air flow and exhalation flow

Oxygen Measurements

Type

Zirconia solid electrolyte

Sensing Position

Internal enclosure

Measurements

Oxygen concentration (range 0 to 95%)

Type

Galvanic cell

Sensing Position

Inspiratory limb of VBS

Measurements

Delivered O

2

% (range 0 to 110%)

Settings, Alarms and
Data Display

Type

TFT liquid crystal touch screen

Table 12-20: Measuring and Display Devices

AC Power & Battery Indicators

Loss of Power (audible only)

Backup Battery Low

External Battery In Use

Backup Battery In Use

Backup Battery Charging

Mains