beautypg.com

Storage, Repairs, Storage -10 repairs -10 – Philips V200 User Manual

Page 132

background image

Chapter 10

10-10

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

Care and Maintenance

Removing and Replacing Internal Air Source Inlet Filter

Figure 10-6: Removing and Replacing Internal Air Source Inlet Filter

Refer to Figure 10-6. Figure 10-6 shows exploded view of the duct (2) and
filter (1), but it does not have to be disassembled.

1. Move the ventilator so that there is easy access to the underside of the

internal air source duct (2).

2. The internal air source filter (1) can be removed manually. The filter

can be reached from under the internal air source duct (2) as
indicated by the arrow.

3. Wash internal air source filter (1) with mild soap and water.

4. Rinse thoroughly then pat dry.

5. Reinstall internal air source filter (1) from underneath the duct (2) in

the same fashion it was removed.

Storage

If you need to store the ventilator for fifteen days or more, ensure that the
altitude, temperature and humidity of the storage site fall within the following
ranges:

Environmental Temperature: -20 to 60°C (-4 to 140° F)

Relative Humidity: 10 to 100% noncondensing

Maximum Altitude: 6,560 m (20,000 ft.)

Repairs

For technical service or repairs not included in this chapter, refer to the Esprit/
V200 Ventilator Service Manual, P/N 580-1000-02 or contact Respironics
Customer Service at 1-800-345-6443.

Pull filter out from
underside of duct.

1

2