beautypg.com

Commands transmitted to the ventilator, Transmission of data from the ventilator, Options and accessories – Philips V200 User Manual

Page 204

background image

Chapter 13

13-40

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

Options and Accessories

Commands
Transmitted to the
Ventilator

Commands are transmitted as a series of four ASCII characters followed by a
carriage return. Valid commands will be stored and response transmissions will
be sent in the order the corresponding commands were received.

Transmission of Data
from the Ventilator

Unless stated otherwise, all fields will be left justified and six (6) characters in
length. A comma will separate each field, except for the field following the
start code. Each data transmission shall be terminated with a carriage return.

In the following subparagraphs, a space is designated as “

◆”. When data is

unavailable, the output field shall contain “

◆◆◆◆◆◆”

SNDA, Send
Variable Length
Ventilator Settings
Report

When the ventilator receives SNDA followed by a carriage return, it will
respond by transmitting the information shown in Table 13-4: “Ventilation
Report”. The ventilator responds to the SNDA command by returning a string
with a variable length. The second and third fields define the length of the
message. The second field indicates the number of characters between the
start and stop codes. The third field indicates the number of fields between
the start and stop codes. The fourth field is the start code, 02H. The last field
in the string is the stop code, 03H, indicating the end of the message.

NOTE:

The SNDA command is only available in Normal mode.

Ventilation Report

Description

Example

Resolution

Range

Units

Comments

Command Name

MISCA

N/A

N/A

N/A

5 character field

Number of characters
between the start and
stop codes

706

N/A

N/A

N/A

3 character field

Number of fields
between the start and
stop codes

97

N/A

N/A

N/A

2 character field

Start Code

0x02

N/A

N/A

N/A

ASCII Start
Transmission
Character (STX)

Time of request

13:45

N/A

N/A

N/A

24 hour clock,
hh:mm

Not used

N/A

N/A

N/A

18 character field.
Outputs example
value.

Table 13-4: Ventilation Report (Sheet 1 of 10)