Rs-232 communications protocol – Philips V200 User Manual
Page 293
Appendix A
RS-232 Communications Protocol
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
Appendix A-7
Measured I:E Ratio
1:4.1
◆
0.1 for 9.9:1
to 1:9.9
1 for 1:10 to
1:99
4.1:1-1:99
N/A
The I:E Ratio
display from the
Patient Data
screen.
Measured Patient
Leak
0.0
◆◆◆
0.1 for 0.0
to 99.9
1 for 100 to
140
0.0 - 140
LPM
The Pt Leak
display.
Measured Percent of
Breaths Triggered by
the Patient (NPPV
Spont/T mode only;
otherwise
◆◆◆◆◆◆)
56.2
0.1 for 0.0
to 99.9
1 for 100
0.0 - 100
%
The Pt Trigger
display on the
NPPV Monitor
screen.
Monitored Oxygen
Concentration
55.1
◆◆
0,1 for 0.0
to 99.9
1 for 100
0.0 - 100
%
The % O2 display
from the Patient
Data Screen.
Ti/Ttot
0.23
◆◆
0.01
0.00 - 1.00
N/A
The Ti/Ttot display
from the
Mechanics Patient
Status Screen
Dynamic Resistance
5.43
◆◆
0.01 for
0.00 to
9.99
0.1 for
10.0 to
99.9
1 for 100
to 400
0.00 - 400
cmH
2
0
/L/Sec
The Dynamic
Resistance display
from the
Mechanics Patient
Status Screen
Dynamic Compliance
19.2
◆◆
0.01 for
0.00 to
9.99
0.1 for
10.0 to
99.9
1 for 100
to 350
0.00 - 350
mL/
cmH
2
0
The Dynamic
Compliance display
from the
Mechanics Patient
Status Screen
Peak Lung Flow
35.1
◆◆
0.1 for -
99.9 to
99.9
1 for -100
to -300
1 for 100
to 300
-300 - 300
LPM
The Peak L-Flow
display from the
Mechanics Patient
Status Screen
Vital Capacity
450
◆◆
1
0 - 9999
mL
Vital Capacity
display from Vital
Capacity Screen
Ventilation Report (Continued)
Description
Example
Resolution
Range
Units
Comments
Table A-1: Ventilation Report (Sheet 6 of 10)