Rs-232 communications protocol – Philips V200 User Manual

Page 293

background image

Appendix A

RS-232 Communications Protocol

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

Appendix A-7

Measured I:E Ratio

1:4.1

0.1 for 9.9:1
to 1:9.9
1 for 1:10 to
1:99

4.1:1-1:99

N/A

The I:E Ratio
display from the
Patient Data
screen.

Measured Patient
Leak

0.0

◆◆◆

0.1 for 0.0
to 99.9
1 for 100 to
140

0.0 - 140

LPM

The Pt Leak
display.

Measured Percent of
Breaths Triggered by
the Patient (NPPV
Spont/T mode only;
otherwise

◆◆◆◆◆◆)

56.2

0.1 for 0.0
to 99.9
1 for 100

0.0 - 100

%

The Pt Trigger
display on the
NPPV Monitor
screen.

Monitored Oxygen
Concentration

55.1

◆◆

0,1 for 0.0
to 99.9
1 for 100

0.0 - 100

%

The % O2 display
from the Patient
Data Screen.

Ti/Ttot

0.23

◆◆

0.01

0.00 - 1.00

N/A

The Ti/Ttot display
from the
Mechanics Patient
Status Screen

Dynamic Resistance

5.43

◆◆

0.01 for
0.00 to
9.99
0.1 for
10.0 to
99.9
1 for 100
to 400

0.00 - 400

cmH

2

0

/L/Sec

The Dynamic
Resistance display
from the
Mechanics Patient
Status Screen

Dynamic Compliance

19.2

◆◆

0.01 for
0.00 to
9.99
0.1 for
10.0 to
99.9
1 for 100
to 350

0.00 - 350

mL/
cmH

2

0

The Dynamic
Compliance display
from the
Mechanics Patient
Status Screen

Peak Lung Flow

35.1

◆◆

0.1 for -
99.9 to
99.9
1 for -100
to -300
1 for 100
to 300

-300 - 300

LPM

The Peak L-Flow
display from the
Mechanics Patient
Status Screen

Vital Capacity

450

◆◆

1

0 - 9999

mL

Vital Capacity
display from Vital
Capacity Screen

Ventilation Report (Continued)

Description

Example

Resolution

Range

Units

Comments

Table A-1: Ventilation Report (Sheet 6 of 10)