Guidance and Manufacturer's Declaration - Electromagnetic Immunity

The ventilator is intended for use in the electromagnetic environment specified below. The user
of the ventilator should assure that it is used in such an environment.

Immunity Test

IEC 60601 Test
Level

Compliance
Level

Electromagnetic Environment -
Guidance

Electrostatic
Discharge (ESD)

IEC 61000-4-2

±6kV contact

±8kV air

±6kV contact

t±8kV air

Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.

Electrical Fast
Transient/burst

IEC 61000-4-4

±2kV for power
supply lines.

±1kV for input/
output lines

±2kV for power
supply lines.

±1kV for input/
output lines

Mains power quality should be
that of a typical hospital
environment

Surge

IEC 61000-4-5

±1kV differential
mode

±2kV common
mode

±1kV differential
mode

±2kV common
mode

Mains power quality should be
that of a typical hospital
environment

Voltage dips, short
interruptions and
voltage variations on
power supply input
lines

IEC 61000-4-11

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

5% U

T

(>95% dip in U

T

)

for 5 sec

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

5% U

T

(>95% dip in U

T

)

for 5 sec

Mains power quality should be
that of a typical hospital
environment. If the user of the
ventilator requires continued
operation during power mains
interruptions, it is recommended
that the ventilator be powered
from an external battery.

Power Frequency
(50/60Hz) magnetic
field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.

Table 12-27: EMC Declaration - Immunity