Appendix A

RS-232 Communications Protocol

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

Appendix A-13

VCV Flow I-Trigger
Setting

3.0

◆◆◆

0.1

0.5 - 20.0

LPM

VCV E-Cycle Type

%Flow

N/A

%FLOW

AUTO

◆◆

N/A

VCV %Flow
Expiratory Cycle
Setting

25

◆◆◆◆

1

10 - 80

%

VCV Rise Time

0.1

◆◆◆

0.1

0.1 - 0.9

Sec

VCV Oxygen
Concentration Setting

21

◆◆◆◆

1

21 - 100

%

VCV Plateau Setting

0.0

◆◆◆

0.1

0.0 - 2.0

Sec

VCV Apnea Rate
Setting

12

◆◆◆◆

1

1 - 80

BPM

VCV High Inspiratory
Pressure Alarm Limit
Setting

35

◆◆◆◆

1

10 - 105

cmH

2

O

(hPa)

VCV Low Inspiratory
Pressure Alarm Limit
Setting

3

◆◆◆◆◆

1

3 - 105

cmH

2

O

(hPa)

VCV Low PEEP Alarm
Limit Setting

0

◆◆◆◆◆

1

0 - 35

cmH

2

O

(hPa)

VCV Low Mandatory
Tidal Volume Alarm
Limit Setting

0

◆◆◆◆◆

1

0 - 2500

mL

VCV Low
Spontaneous Tidal
Volume Alarm Limit
Setting

0

◆◆◆◆◆

1

0 - 2500

mL

VCV High Respiratory
Rate Alarm Limit
Setting

150

◆◆◆

1

0 - 150

BPM

VCV Low Minute
Volume Alarm Limit
Setting

1.00

◆◆◆

0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0

0.00 - 60.0

L

VCV Apnea Interval
Alarm Limit Setting

20

◆◆◆◆

1

10 - 60

sec

VCV High Minute
Volume Alarm Limit
Setting

60.0

◆◆

0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0

0.00 - 60.0

L

Stop Code

0x03

N/A

N/A

N/A

ASCII End
Transmission
Character (ETX)

Volume Control Ventilation Settings Report (Continued)

Description

Example

Resolution

Range

Units

Comments

Table A-2: Volume Control Ventilation Settings Report (Sheet 2 of 2)