Alarms, Alarms -86, Options and accessories – Philips V200 User Manual
Page 250
Chapter 13
13-86
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Table 13-14: Respiratory Mechanics Maneuvers with Flow-Trak
Alarms
The Flow-Trak option does not change the ventilator alarms with the exception
of those alarms found below in Table 13-15:. For a complete listing of
ventilator alarms and descriptions see Chapter 9, “Alarms”, Table 9-1: “Alarm
Alert Messages”.
Table 13-15: Flow-Trak Affected Alarms
NOTE:
The I-Time too long alarm and/or the Low Inspiratory Pressure alarm will
sound if a circuit disconnect were to occur.
Respiratory Mechanics Maneuvers with Flow-Trak
Maneuver
Affect of Flow-Trak
Static C & R
The maneuver is cancelled if the patient initiates a Flow-Trak breath. A
message appears when the maneuver is cancelled due to the change in breath
type.
Dynamic C & R
Dynamic C & R are not calculated for Flow-Trak breaths.
Ti/Ttot
Ti/Ttot is not calculated for Flow-Trak breaths.
Peak Lung Flow
Peak Lung Flow is calculated for Flow-Trak breaths.
Flow-Trak Affected Alarms
Alert Message
Description
I-Time too long
The breath transitions from inhalation to exhalation when 3.5
seconds has passed for adults or 2.5 seconds for pediatrics from
the start of the mandatory breath. If the operator’s settings result
in a calculated I-Time > 3.5 seconds for adults or 2.5 seconds for
pediatrics, then the calculated I-Time is the cycling criteria.
Low Inspiratory Pressure
Not available for the Flow-Trak triggered breath only.