Technical Specifications

Front Panel Keys

I:E

9.9:1 to 1:99

none

0.1 for 9.9:1-1:9.9;
1 for 1:10-1:99

±10%

Peak Inspiration
Pressure

-20.0 to 130

cmH

2

O

(hPa)

1 for 100-130;
0.1 for -20.0 to 99.9

±10%

End Inspiration
Pressure (Pi End)

-20.0 to 130

cmH

2

O

(hPa)

1 for 100-130;
0.1 for -20.0 to 99.9

±10%

Mean Airway
Pressure

-20.0 to 120

cmH

2

O

(hPa)

1 for 100-120;
0.1 for -20.0 to 99.9

±10%

Delivered O

2

%

0.0 to 110

--

1 for 100-110;
0.1 for -20.0 to 99.9

±3 vol%

Total Respiratory Rate

0.0 to 150

Bpm

1 for 10-150;
0.1 for 0.0 to 9.9

±10%

Spontaneous
Respiratory Rate

0.0 to 150

Bpm

1 for 10-150;
0.1 for 0.0 to 9.9

±10%

End Exhalation
Pressure (Pe End)

-20.0 to 99.9

cmH

2

O

(hPa)

0.1

±10%

Patient Leak (Pt Leak)

0.0 to 140

LPM

0.1 for 0.0-99.9
1 for 100-140

Not
specified

a

Percent Patient Trigger
(%Pt Trigger)

0.0 to 99.9

%

0.1 for 0.0-99.9
1 for 100

±10%

a. Due to the variable nature of patient leaks, this parameter is an estimate only.

Patient Data Range, Resolution, Units & Accuracy (Continued)

Patient Data

Display Range

Units

Resolution

Accuracy

Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 2 of 2)

Front Panel Keys

Alarm Silence

Screen Lock

Alarm Reset

Expiratory Hold

100% O

2

Accept

Manual Breath

Table 12-11: Front Panel Keys